Select a Zero Method
SWIFT 10 MC Sensors Setting up the Transducer Interface
51
Calibration File Elements
The following figure shows some elements of the calibration file:
Typical Calibration File
Items you may edit OutputPolarities—defines the polarities of the six outputs. Change these
only if your application requires different polarities from those identified on
the transducer label.
AngleMode—selects the mode used for determining the encoder sine and
cosine.
AngleFixed—used for non-spinning applications.
AngleOffset— used for spinning applications. Normally you do not need to
change this value.
EncoderSize–defines the size of the encoder.
S10MC-33
SerialNum=10237221A
AngleMode=0
AngleOffset=0
AngleFixed=0
EncoderSize=2048
FXPolarity=0
FYPolarity=0
FZPolarity=0
MXPolarity=1
MYPolarity=0
MZPolarity=1
ZFX1=0
ZFX2=0
ZFY1=0
ZFY2=0
ZFY3=0
ZFY4=0
ZFZ1=0
ZFZ2=0
FX1Channel=2
FX2Channel=6
FY1Channel=99
FY2Channel=99
FY3Channel=99
FY4Channel=99
FZ1Channel=4
FZ2Channel=0
InputRange=10.0
KFX1=0.492315
KFX2=0.492315
KFY1=0.0
KFY2=0.0
KFY3=0.0
KFY4=0.0
KFZ1=0.49212
KFZ2=0.49212
KMX=0.0
KMYX=1.44278
KMYZ=1.44278
KMZ=0.0
KFXFY=0.0
Polarity for each output
(0 = normal, 1 = inverted)
Angle Mode (0 = Spinning, 1 = Fixed)
Angle Offset
Fixed Angle
Bridge Zeroes
Do Not Modify
Calibration Gains
Do Not Modify
Encoder Size