J9 Interlock OUT
Models FlexTest® IIm/GT/SE Controller Hardware FlexTest GT Controller Connections
173
Cable specifications Connector: 25-contact, type D, male, EMI
Backshell: EMI, metalized plastic
Twisted pairs with overall braided shield. Braided shield is terminated to
conductive backshell at the chassis.
Jumper plug Use MTS jumper plug PN 100-079-233.
J9 Interlock OUT
Connector J9 Interlock OUT provides four station interlock outputs that can be
connected to a multiple MTS controller box interlock chain.
Electrical
specifications
Normally, Open = Interlock, 1 A maximum @ 30 V DC/AC maximum
Cable specification Connector: 9-contact, type D, male, EMI
Backshell: EMI metalized plastic
Twisted pairs with overall braided shield. Braided shield is terminated to
conductive backshell at the chassis.
+
-
Interlock Status
Output 2
+
-
Interlock Status
Output 1
+
-
Interlock Status
Output 3
+
-
Interlock Status
Output 4
1
6
2
7
3
8
4
9
J9 To
External Devices