Mts FlexTest Models 200 Uživatelský manuál Strana 291

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 344
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 290
Optional Station Configurations
MTS FlexTest® Models 40/60/100/200 Controller Hardware Optional Station Configurations
291
Appendix B
Optional Station Configurations
About Optional Station Configurations
You can configure the Series 494 Chassis to support an optional six or eight
stations. When configuring your system for a six or eight stations, consider
the following:
Ensure that the .hwi file is correctly set for the desired multistation
configuration, especially the interlock and HSM board settings.
Power to each HSM is limited.
Cross-head interlocks with solenoid power are not supported.
AC input power must be at least 115 V AC.
Auxiliary power out of J49 on the Model 493.74 HSM Transition board
is not supported.
Six-Station Configurations
The six-station configuration can provide either six or eight channels of
control.
A typical 6-channel/6-station configuration requires 12 Digital
Universal Conditioners and 6 two-stage valve drivers.
A typical 8-channel/6-station configuration requires 16 Digital
Universal Conditioners and 8 two-stage valve drivers.
HSM power limits HSM power current is limited to 1.6 A per HSM.
Interlocks For 6-station configurations, the .hwi file must contain the line
INTERLOCKS=6. This line must be a discrete entry, not part of any other
hwi section.
Cross-head interlocks with solenoid power are not supported for 6-station
configurations.
Zobrazit stránku 290
1 2 ... 286 287 288 289 290 291 292 293 294 295 296 ... 343 344

Komentáře k této Příručce

Žádné komentáře