Series 609 Alignment Fixture Product Information10ConventionsPrefaceConventionsDocumentation ConventionsThe following paragraphs describe some of the
Documentation ConventionsSeries 609 Alignment Fixture Product Information Preface11Hypertext links The electronic document has many hypertext links di
Series 609 Alignment Fixture Product Information12Documentation ConventionsPreface
Series 609 Alignment Fixture Product Information Introduction13IntroductionThe MTS Series 609 Alignment Fixture improves alignment between the upper a
Series 609 Alignment Fixture Product Information14IntroductionWhat you need toknowMTS Systems Corporation assumes that you know how to use your contro
MisalignmentSeries 609 Alignment Fixture Product Information Introduction15MisalignmentIn uniaxial testing, perfectly aligned grips produce uniform ax
Series 609 Alignment Fixture Product Information16Adjust Concentric AlignmentIntroductiongage section. These higher strains are on opposite sides. It
About Angular MisalignmentSeries 609 Alignment Fixture Product Information Introduction17Improving Concentric AlignmentAbout Angular MisalignmentAngul
Series 609 Alignment Fixture Product Information18Adjust Angular AlignmentIntroductionAngular Misalignment Produces a “C” BendAdjust Angular Alignment
Adjust Angular AlignmentSeries 609 Alignment Fixture Product Information Introduction19Improving Angular AlignmentHousingAngularity CollarAngularity A
Manual Template 4.32Copyright information © 1993, 1999, 2001, 2008 MTS Systems Corporation. All rights reserved.Trademark information MTS is a registe
Series 609 Alignment Fixture Product Information20Adjust Angular AlignmentIntroduction
Hazard Placard PlacementSeries 609 Alignment Fixture Product Information Safety Information21Safety InformationHazard Placard PlacementHazard placards
Series 609 Alignment Fixture Product Information22Hazard Placard PlacementSafety Information
Series 609 Alignment Fixture Product Information Installation23InstallationThis section describes how to install the Series 609 Alignment Fixture. As
Series 609 Alignment Fixture Product Information24InstallationNote The following procedure assumes the force transducer is mounted to the crosshead. I
Series 609 Alignment Fixture Product Information Installation25For the jackbolts, lightly lubricate the preload collar and its washer with Molykote G-
Series 609 Alignment Fixture Product Information26InstallationE. Install the force transducer and alignment fixture assembly into the crosshead. Caref
Series 609 Alignment Fixture Product Information Installation27To reduce the hazards in this procedure:• Ensure that you set and enable displacement i
Series 609 Alignment Fixture Product Information28InstallationF. If the TIR is 0.038 mm (0.0015 in) or less, the force transducer is accurately aligne
About Gaged SpecimensSeries 609 Alignment Fixture Product Information Specimen Preparation29Specimen PreparationThis section shows you where to place
Series 609 Alignment Fixture Product Information Contents3ContentsTechnical Support 5How to Get Technical Support 5Before You Contact MTS 5If
Series 609 Alignment Fixture Product Information30About Gaged SpecimensSpecimen PreparationBecause misalignment produces bending strain, many test pro
Round Thick Diameter SpecimensSeries 609 Alignment Fixture Product Information Specimen Preparation31Round Thick Diameter SpecimensA typical gaged spe
Series 609 Alignment Fixture Product Information32Calculating Bending Strain—Round Thick Diameter Specimen PreparationCalculating Bending Strain—Round
Calculating Bending Strain—Round Thick DiameterSeries 609 Alignment Fixture Product Information Specimen Preparation33Use this formula to find the ave
Series 609 Alignment Fixture Product Information34Round Thin Diameter SpecimensSpecimen PreparationRound Thin Diameter SpecimensA typical gaged specim
Calculating Bending Strain—Round Thin DiameterSeries 609 Alignment Fixture Product Information Specimen Preparation35Calculating Bending Strain—Round
Series 609 Alignment Fixture Product Information36Calculating Bending Strain—Round Thin Diameter Specimen Preparation4. Calculate the direction of the
Flat Thick SpecimensSeries 609 Alignment Fixture Product Information Specimen Preparation37Flat Thick SpecimensA typical gaged specimen has three leve
Series 609 Alignment Fixture Product Information38Calculating Bending Strain—Flat Thick SpecimensSpecimen PreparationCalculating Bending Strain—Flat T
Calculating Bending Strain—Flat Thick SpecimensSeries 609 Alignment Fixture Product Information Specimen Preparation393. Calculate the percent bending
Series 609 Alignment Fixture Product Information4ContentsSpecimen Preparation 29About Gaged Specimens 29Round Thick Diameter Specimens 31Calc
Series 609 Alignment Fixture Product Information40Flat Thin SpecimensSpecimen PreparationFlat Thin SpecimensA typical gaged specimen has three levels
Calculating Bending Strain—Flat Thin SpecimensSeries 609 Alignment Fixture Product Information Specimen Preparation41Calculating Bending Strain—Flat T
Series 609 Alignment Fixture Product Information42Notched Round Thick Diameter SpecimensSpecimen PreparationWhere: See the definitions above.Notched R
Calculating Bending Strain—Notched Round ThickSeries 609 Alignment Fixture Product Information Specimen Preparation43Calculating Bending Strain—Notche
Series 609 Alignment Fixture Product Information44Notched Round Thin Diameter SpecimensSpecimen PreparationWhere: ε1 through ε8 are the strain reading
Calculating Bending Strain—Notched Round ThinSeries 609 Alignment Fixture Product Information Specimen Preparation45Gage Placement—Notched Three Gage
Series 609 Alignment Fixture Product Information46Calculating Bending Strain—Notched Round Thin Specimen Preparation1. Calculate the average axial str
Calculating Bending Strain—Notched Round ThinSeries 609 Alignment Fixture Product Information Specimen Preparation47Where: (φ) is measured from the hi
Series 609 Alignment Fixture Product Information48Notched Flat Thick SpecimensSpecimen PreparationNotched Flat Thick SpecimensA typical gaged specimen
Calculating Bending Strain—Notched Flat ThickSeries 609 Alignment Fixture Product Information Specimen Preparation49Calculating Bending Strain—Notched
How to Get Technical SupportSeries 609 Alignment Fixture Product Information Technical Support5Technical SupportHow to Get Technical SupportStart with
Series 609 Alignment Fixture Product Information50Calculating Bending Strain—Notched Flat Thick Specimen PreparationWhere: Bu is the bending strain of
Notched Thin Flat SpecimensSeries 609 Alignment Fixture Product Information Specimen Preparation51Notched Thin Flat SpecimensA notched gaged specimen
Series 609 Alignment Fixture Product Information52Calculating Bending Strain—Notched Thin Flat Specimen PreparationCalculating Bending Strain—Notched
Calculating Bending Strain—Notched Thin FlatSeries 609 Alignment Fixture Product Information Specimen Preparation53• Bending strain along one axis (us
Series 609 Alignment Fixture Product Information54Calculating Bending Strain—Notched Thin Flat Specimen Preparation
Series 609 Alignment Fixture Product Information Alignment Procedure55Alignment ProcedureThis section covers alignment using the two most common gaged
Series 609 Alignment Fixture Product Information56Installing a Gaged SpecimenAlignment ProcedureInstalling a Gaged SpecimenIn this section, you clamp
Installing a Gaged SpecimenSeries 609 Alignment Fixture Product Information Alignment Procedure57Select force control as your active control mode. Sel
Series 609 Alignment Fixture Product Information58Adjust the Concentric AlignmentAlignment ProcedureAdjust the Concentric AlignmentIn this section, yo
Adjust the Concentric AlignmentSeries 609 Alignment Fixture Product Information Alignment Procedure593. Using the following figure, determine which sc
Series 609 Alignment Fixture Product Information6If You Contact MTS by PhoneTechnical SupportKnow information fromprior technicalassistanceIf you have
Series 609 Alignment Fixture Product Information60Adjust the Concentric AlignmentAlignment ProcedureRound Specimens—Reduce the “S” BendRound specimens
Adjust the Concentric AlignmentSeries 609 Alignment Fixture Product Information Alignment Procedure61The following figure shows typical readings befor
Series 609 Alignment Fixture Product Information62Adjust the Concentric AlignmentAlignment Procedure5. Continue to adjust concentricity screws to redu
Adjust the Angular AlignmentSeries 609 Alignment Fixture Product Information Alignment Procedure63Adjust the Angular AlignmentAn angular adjustment ti
Series 609 Alignment Fixture Product Information64Adjust the Angular AlignmentAlignment Procedure2. Using the following figure, determine which screws
Adjust the Angular AlignmentSeries 609 Alignment Fixture Product Information Alignment Procedure65Round Specimens—Reduce the “C” BendIf the magnitude
Series 609 Alignment Fixture Product Information66Adjust the Angular AlignmentAlignment ProcedureThe following figure shows typical readings before an
Check the AlignmentSeries 609 Alignment Fixture Product Information Alignment Procedure67Find Each Level’s Maximum Bending Strain and Direction3. Calc
Series 609 Alignment Fixture Product Information68Check the AlignmentAlignment ProcedureYour gaged specimen may not be strong enough to handle the ful
Problem Submittal Form in MTS ManualsSeries 609 Alignment Fixture Product Information Technical Support7If you are calling about an issue that has alr